Thermo Scientific Orbitrap Fusion / UltiMate 3000 nanoLC System jpg

Thermo Scientific Orbitrap Fusion / UltiMate 3000 nanoLC System

Equipment overview

The Thermo Scientific Orbitrap Fusion Liquid Chromatography/Tandem Mass Spectrometer is a high-resolution, accurate-mass (HRAM) multi-attribute tribrid mass spectrometer designed to deliver exceptional performance, operational versatility, and robust quantification and qualitative capabilities for complex proteomics, small molecule analysis, and structural elucidation.

Major applications

Proteomics (global, quantitative, and PTM profiling), metabolomics, lipidomics, macromolecular complex analysis, and small molecule structure elucidation.

Specifications

Liquid Chromatography
Thermo Scientific UltiMate 3000 RSLCnano System
• Mass Spectrometer
Thermo Scientific Orbitrap Fusion Tribrid Mass Spectrometer
• Ionization Modes
Electrospray Ionization (ESI) and Nano-Electrospray Ionization (NSI)
• Configuration
Quadrupole-Linear Ion Trap-Orbitrap
• Acquisition Modes
Full MS Scan, Product Ion Scan (MS2/MSn), Data-Dependent Acquisition (DDA), Data-Independent Acquisition (DIA), targeted SIM (tSIM), targeted MS2 (tMS2), and Multiple Dissociation Techniques (CID, HCD, ETD)
• Mass Range
m/z 50–2,000 (standard); m/z 200–4,000 (high mass); up to m/z 6,000 without precursor ion selection
• Mass Resolution
Up to 450,000 (FWHM) at m/z 200
• Maximum Scan Speed
Up to 20 Hz (Orbitrap)
• Mass Accuracy
<3 ppm RMS using external calibration; <1 ppm RMS using internal calibration.
• Software
Orbitrap Fusion Instrument Control Software, Chromeleon CDS, Xcalibur